Titel Comparison of photometer calibrations at six different facilities of PTB and NIST
Autor(in); Institution
Winter, Stefan; 4.1, Photometrie und angewandte Radiometrie, PTB-Braunschweig
Lindner, Detlef; 4.1, Photometrie und angewandte Radiometrie, PTB-Braunschweig
Sperling, Armin; 4.1, Photometrie und angewandte Radiometrie, PTB-Braunschweig
Sauter, Georg; 4.1, Photometrie und angewandte Radiometrie, PTB-Braunschweig
Brown, Steven; NIST, Gaithersburg, USA
Larason, Thomas; NIST, Gaithersburg, USA
Zong, Y.; NIST, Gaithersburg, USA
Ohno, Yoshi; NIST, Gaithersburg, USA
Quelle/Jahr Proceedings of the 9th International Conference on New Developments and Applications in Optical Radiometry:(2005), 243 - 244
Herausgeber(in)
Gröbner, Julian
ISBN 3-03-300570-5 ; 978-3-033-00570-9
Verlag Davos: WRC PMOD
Konferenzangaben 9th International Conference on New Developments and Applications in Optical Radiometry, Davos, 17-19, October, 2005, Switzerland
Zusammenfassung Two photometers were calibrated at six different facilities of the PTB and the NIST. At each institute a secondary calibration against other reference photometers at an optical bench, a primary calibration against reference photodiodes at monochromator facilities and a primary calibration with tunable lasers were performed. All measurements are traceable to the respective cryogenic radiometers used as primary and national standards. Most of the measurements have been completed; the remaining measurements will be done this summer before NEWRAD. The deviations of the already performed measurements are within ± 0.6 %.

Zitierung

Winter, S., Lindner, D., Sperling, A., Sauter, G., Brown, S., Larason, T., Zong, Y., & Ohno, Y. (2005). Comparison of photometer calibrations at six different facilities of PTB and NIST. 9th International Conference on New Developments and Applications in Optical Radiometry, Davos, 17-19, October, 2005, Switzerland.