Zusammenfassung This paper describes the set-up and calibration of an inductive bridge standard which realizes selected voltgae ratios - in particular between 0.1 and 1.3 - at low frequencies of a few 10 Hz with uncertainties around 3·10-8 (k = 2). This allows ac thermometry bridges to be traceably calibrated with uncertainties below 5·10-8.

Zitierung

Ramm, G., Moser, H., & Vollmert, R. (2006). Traceable inductive bridge standard for calibrating ac resistance thermometry bridges at low frequencies [poster]. CPEM 2006, Conference on Precision Electromagnetic Measurements, Torino, 9-14, July, 2006, Italy.