Zugriffsnummer 19015
Dokumenttyp Konferenzartikel in Zeitschrift
Sprache Englisch
Titel Quantitative linewidth measurement down to 100 nm by means of optical dark-field microscopy and rigorous model-based evaluation
Autor(in); Institution
Ehret, Gerd; 4.2, Bild- und Wellenoptik, PTB-Braunschweig
Bodermann, Bernd; 4.2, Bild- und Wellenoptik, PTB-Braunschweig
Mirandé, Werner; 4.2, Bild- und Wellenoptik, PTB-Braunschweig
Quelle/Jahr Nanoscale metrology. Measurement Science and Technology: 18 (2007), 2, 430 - 438
ISSN 0957-0233
DOI
URL
Verlag Bristol: Institute of Physics (IOP)
Konferenzangaben 7th Seminar on Quantitative Microscopy (QM) and 3rd Seminar on Nanoscale Calibration Standards and Methods, Nanoscale 2006, Bern-Wabern, 24-25, April, 2006, Switzerland
Freie Schlagworte darkfield microscopy ; optical CD measurement ; rigorous grating diffraction
Zusammenfassung he dark-field microscopy method with alternating grazing incidence UV illumination (UV-AGID) developed at the PTB now offers the possibility of measuring lateral structures down to 100 nm. The advantage of this new dark-field method is, in comparison to bright-field microscopy, the better edge localization, since only light that is scattered or diffracted by object details reaches the image plane. Also, in the case of AGID microscopy, the proximity effects are much more reduced. This method can be used for dimensional measurements of micro- and nanostructures, e.g. on wafers, masks or etched Si structures. Determining the linewidth quantitatively from the microscope image is, however, only possible by comparing the measured data with adapted rigorous simulations. To model the intensity distribution in the image, we make use of two different rigorous grating diffraction theories. On the one hand we use the rigorous coupled wave analysis (RCWA) method (for calculations of the diffracted electrical and magnetic fields), on the other hand, the finite elements method (FEM). In particular, it will be discussed what influence the polarization of the illumination and the edge angles of the trapezoidal-shaped lines and spaces have on the microscope image. Also, different quantitative measurements of chromium lines on a COG mask (COG: chromium structures on glass) will be presented. The results will be compared with electron optical measurements.

Zitierung

Ehret, G., Bodermann, B., & Mirandé, W. (2007). Quantitative linewidth measurement down to 100 nm by means of optical dark-field microscopy and rigorous model-based evaluation. Measurement Science and Technology, 18(2), 430–438. https://doi.org/10.1088/0957-0233/18/2/s15

Exportieren

PTB-Publica Menü

Sprache wechseln: uk flag