| Titel | A novel method of radiochromic film dosimetry using a color scanner |
| Autor(in); Institution |
Hupe, Oliver; 6.3, Strahlenschutzdosimetrie, PTB-Braunschweig
Brunzendorf, Jens; 6.3, Strahlenschutzdosimetrie, PTB-Braunschweig
|
| Quelle/Jahr | Medical Physics: 33 (2006), 11, 4085 - 4094 |
| ISSN | 0094-2405 |
| DOI | |
| Verlag | New York, NY: AAPM |
| Klassifikationscode | PACS: 87.66.Cd ; PACS: 87.66.Jj ; PACS: 87.53.Jw ; PACS: 87.58.Sp ; PACS: 87.53.Xd |
| Freie Schlagworte | brachytherapy ; calibration ; dosimetry ; film digitization ; film dosimetry ; film scanners ; GAFChromic ; quality assurance ; quality control ; radiochromic ; signal-to-noise ratio ; homogeneity ; HS ; MD55 ; EBT |
| Zusammenfassung | A procedure that allows the improved extraction of the dose information based on the multi-colour scanning of the radiochromic film is presented. The basic principle is the determination of the dose values from each colour channel of the digital film image in RGB-format by applying a non-linear calibration function. The best estimate of the dose is then a weighted mean of the dose values derived from each colour channel. The weighting factors are determined in such a way that the noise in the 2-dimensional dose profile is a minimum. The calculation of the weighting factors is presented, they are chosen to be proportional to the signal-to-noise ratio, Si /νi2 , in all three colour channels, i= red, green or blue. The data reduction can be fully com-puterized, including the cleaning of the digital image from dust and scratches. It is highly reproducible, which is important for quality assurance, and easy to use. Our novel evaluation procedure combines the good response in the low dose range of the red colour with the extended dose range of the blue and green colour channels (response up to 10000 Gy) making use of one single steady evaluation function. Therefore, a smooth evaluation is possible in a wide dose range. For the interpretation of measurements with the radiochromic films the spatial inhomogeneity of the film's response to ionizing radiation is very important. Investigations on both film types, HS and MD55-2, as well the new EBT film have been carried out. |