| Zugriffsnummer | 18297 |
| Dokumenttyp | Konferenzartikel |
| Sprache | Englisch |
| Titel | Determination of the unit of the inductance with the Maxwell-Wien-bridge in PTB |
| Autor(in); Institution |
Hanke, Rüdiger; 2.1, Gleichstrom und Niederfrequenz, PTB-Braunschweig
Skubis, Tadeusz; Silesian Technical University of Technology, Gliwice, POLAND
Kölling, Axel; 2.1, Gleichstrom und Niederfrequenz, PTB-Braunschweig
|
| Quelle/Jahr | [Regional Metrology Conference - RMC 2005]:(2005) |
| Availability | [CD-ROM] ; file name: Nairobi_Hanke.doc |
| Verlag | s.l.: |
| Konferenzangaben | Regional Metrology Conference 2005, Nairobi, 19-20, September, 2005, Kenya |
| Zusammenfassung | For the determination of the unit Henry of the inductance at PTB a Maxwell-Wien bridge has been developed. The advantage of this type of bridge is that the first order equation is independent of frequency and therefore works well in a frequency range from 50 Hz up to 1MHz. The bridge includes a Wagner device to eliminate stray and cable capacitances. It is necessary to consider the frequency dependence of the used primary standards to decrease the desired uncertainties of the determinated unit Henry. |