Zugriffsnummer 18112
Dokumenttyp Zeitschriftenartikel
Sprache Englisch
Titel Response of a BaF2 scintillation detector to quasi-monoenergetic fast neutrons in the range of 45 to 198 MeV
Autor(in); Institution
Gunzert-Marx, K.; Gesellschaft für Schwerionenforschung, Darmstadt, GERMANY
Schardt, D.; Gesellschaft für Schwerionenforschung, Darmstadt, GERMANY
Simon, R. S.; Gesellschaft für Schwerionenforschung, Darmstadt, GERMANY
Gutermuth, F.; Gesellschaft für Schwerionenforschung, Darmstadt, GERMANY
Radon, T.; Gesellschaft für Schwerionenforschung, Darmstadt, GERMANY
Dangendorf, Volker; 6.5, Neutronenstrahlung, PTB-Braunschweig
Nolte, Ralf; 6.4, Ionenbeschleuniger und Referenzstrahlungsfelder, PTB-Braunschweig
Quelle/Jahr Nuclear Instruments and Methods in Physics Research A:(2005), 536, 146 - 153
ISSN 0168-9002
DOI
Verlag Elsevier
Freie Schlagworte Bariumfluoride scintillator ; Neutron efficiency ; Quasi-monoenergetic neutron beam ; Monte Carlo code FLUKA
Zusammenfassung We have studied the neutron response of a scintillation detector consisting of a 14 cm long, hexagonal-shaped BaF2-crystal with an inner diameter of 8.75 cm coupled to an EMI9821QB photomultiplier tube. The detector was exposed to calibrated quasi-monoenergetic neutron fields obtained from 7Li(p,n)7Be reactions. The measurements were performed at neutron energies of 45, 60, 96, 147 and 198 MeV as given by the energies of the incident protons. The experimental pulse-height spectra of the BaF2-detector are compared with Monte Carlo simulations using the FLUKA code. The detection efficiency of the BaF2-detector in the energy range of 45-198 MeV was determined as a function of the discriminator threshold and compared to the literature data. At neutron energies above 100 MeV the detection efficiency of the BaF2-detector was found to be a factor of two higher than that of an NE213-detector comparable size.

Zitierung

Gunzert-Marx, K., Schardt, D., Simon, R. S., Gutermuth, F., Radon, T., Dangendorf, V., & Nolte, R. (2005). Response of a BaF₂ scintillation detector to quasi-monoenergetic fast neutrons in the range of 45 to 198 MeV. Nuclear Instruments and Methods in Physics Research A, 146–153. https://doi.org/10.1016/j.nima.2004.07.168

Exportieren