| Zugriffsnummer | 17978 |
| Dokumenttyp | Konferenzartikel in Zeitschrift |
| Peer Review | unbekannt |
| Sprache | Englisch |
| Titel | New primary standard for the realization of reference air kerma rate for low-dose rate, low-energy brachytherapy sources |
| Autor(in); Institution |
Selbach, Hans-Joachim; 6.2, Dosimetrie für die Strahlentherapie, PTB-Braunschweig
Culberson, Wesley Stuart; University of Wisconsin-Madison, USA
|
| Quelle/Jahr | Medical physics : proceedings of the jointly held congresses ICMP 2005 .... Biomedizinische Technik: 50 (2005), Suppl. 1, Pt.1, 764 - 765 |
| ISSN | 0939-4990 |
| Verlag | Berlin: Schiele & Schön |
| Konferenzangaben | ICMP 2005 - 14. International Conference of Medical Physics of the International Organization for Medical Physics (IOMP), the European Federation of Organizations in Medical Physics (EFOMP) and the German Society of Medical Physics (DGMP), Nürnberg, 14-17, September, 2005, Germany |
| Freie Schlagworte | reference air-kerma rate ; brachytherapy ; prostate seed calibration |
| Zusammenfassung | A large air-filled parallel-plate extrapolation chamber (GROVEX) with thin graphite front and back electrodes is used as a primary standard measuring device for low-energy photon brachytherapy sources from which the unit of air kerma strength or reference air kerma can be derived. The setup of the extrapolation chamber is described together with the method of evaluation of the extrapolation curves. Initial air kerma rate measurements with this chamber of previously NIST calibrated 125I seeds are presented. |