Zugriffsnummer 17813
Dokumenttyp Zeitschriftenartikel
Sprache Englisch
Titel Solar-blind diamond detectors for LYRA, the solar VUV radiometer on board PROBA II
Autor(in); Institution
Benmoussa, A.; Royal Observatory of Belgium, Brussels, BELGIUM
Hochedez, J.-F.; Royal Observatory of Belgium, Brussels, BELGIUM
Schmutz, W.K.; Physikalisch-Meteorologisches Observatorium Davos - World Radiation Center, Davos Dorf, SWITZERLAND
Schühle, U.; Max Planck Institut für Sonnensystemforschung, Katlenburg-Lindau, GERMANY
Nesladék, M.; Institute for Materials Research (IMO), Limburgs Universitaire Centrum, Diepenbeek, Belgium, Division IMOMEC, IMEC vzw, Diepenbeek, BELGIUM
Stockmann, Y.; Centre Spatial de Liège, Angleur, BELGIUM
Kroth, Udo; 7.1, Photonenradiometrie, PTB-Berlin
Richter, Mathias; 7.1, Photonenradiometrie, PTB-Berlin
Theissen, A.; Royal Observatory of Belgium, Brussels, BELGIUM
Remes, Z.; Institute for Materials Research (IMO), Limburgs Universitaire Centrum, Diepenbeek, BELGIUM
Haenen, K.; Institute for Materials Research (IMO), Limburgs Universitaire Centrum, Diepenbeek, Belgium, Division IMOMEC, IMEC vzw, Diepenbeek, BELGIUM
Mortet, V.; Institute for Materials Research (IMO), Limburgs Universitaire Centrum, Diepenbeek, BELGIUM
Koller, S.; Physikalisch-Meteorologisches Observatorium Davos - World Radiation Center, Davos Dorf, SWITZERLAND
Halain, J.P.; Centre Spatial de Liège, Angleur, BELGIUM
Petersen, R.; Division IMOMEC, IMEC vzw, Diepenbeek, BELGIUM
Dominique, M.; Royal Observatory of Belgium, Brussels, BELGIUM
D'Olieslaeger, M.; Institute for Materials Research (IMO), Limburgs Universitaire Centrum, Diepenbeek, Belgium, Division IMOMEC, IMEC vzw, Diepenbeek, BELGIUM
Quelle/Jahr Experimental Astronomy: 16 (2003), 141 - 148
ISSN 0922-6435
DOI
Verlag Dordrecht [u.a.]: Springer Science
Freie Schlagworte diamond ; solar-blind photodetector ; UV detectors
Zusammenfassung Fabrication, packaging and experimental results on the calibration of metal-semiconductormetal (MSM) photodetectors made on diamond are reported. LYRA (Lyman-α RAdiometer onboard PROBA-2) will use diamond detectors for the first time in space for a solar physics instrument. A set of measurement campaigns was designed to obtain the XUV-to-VIS responsivity of the devices and other characterizations. The measurements of responsivity in EUV and VUV spectral ranges (40–240 nm) have been carried out by the Physikalisch-Technische Bundesanstalt (PTB) in Germany at the electron storage ring BESSY II. The longer wavelength range from 210 to 1127 nm was measured with monochromatic light by using a Xe-lamp at IMO-IMOMEC. The diamond detectors exhibit a photoresponse which lie in the 35–65 mA/W range at 200 nm (corresponding to an external quantum efficiency of 20–40%) and indicate a visible rejection ratio (200–500 nm) higher than four orders of magnitude.

Zitierung

Benmoussa, A., Hochedez, J.-F., Schmutz, W., Schühle, U., Nesladék, M., Stockmann, Y., Kroth, U., Richter, M., Theissen, A., Remes, Z., Haenen, K., Mortet, V., Koller, S., Halain, J., Petersen, R., Dominique, M., & D'Olieslaeger, M. (2003). Solar-blind diamond detectors for LYRA, the solar VUV radiometer on board PROBA II. Experimental Astronomy, 16, 141–148. https://doi.org/10.1007/s10686-003-0035-3

Exportieren

PTB-Publica Menü