Titel APMP international comparison of ac-dc transfer standards at the lowest attainable level of uncertainty
Autor(in); Institution
Budovsky, Ilya; National Measurement Institute of Australia (NMIA), Lindfield, AUSTRALIA
Bin Zainal Abidin, Abdul Rashid; National Metrology Laboratory (SIRIM), Shah Alam, MALAYSIA
Yan, Aaron Y. K.; Hong Kong Standards and Calibration Laboratory (SCL), Wanchai, CHINA
Liu, LingXiang; Standards, Productivity, and Innovation Board (SPRING), SINGAPORE
Rustagi, Vijay Kumar; National Physical Laboratory (NPLI), New Delhi, INDIA
Govil, Anil Kumar; National Physical Laboratory (NPLI), New Delhi, INDIA
Klonz, Manfred; 2.1, Gleichstrom und Niederfrequenz, PTB-Braunschweig
Wei, Yih-cheng; Center for Measurement Standards, Industrial Technology Research Institute (ITRI), Hsinchu, TAIWAN
Early, Murray D.; Measurement Standards Laboratory (MSL), Lower Hutt, NEW ZEALAND
Sasaki, Hitoshi; National Metrology Institute of Japan (NMIJ), AIST, Tsukuba, JAPAN
Fujiki, Hiroyuki; National Metrology Institute of Japan (NMIJ), AIST, Tsukuba, JAPAN
Kumtawee, Chalit; National Institute of Metrology, Thailand (NIMT), Bangkok, THAILAND
Charoensook, Ajchara; National Institute of Metrology, Thailand (NIMT), Bangkok, THAILAND
Kwon, Sung-Won; Korea Research Institute of Standards and Science (KRISS), Daejon, KOREA
Son, Nguyen Anh; Vietnam Metrology Institute (VMI), Hanoi, VIETNAM
Suprianto, Bambang; Pusat Penelitian Kalibrasi, Instrumentasi, dan Metrologi (KIM-LIPI), Jakarta, INDONESIA
Temba, Moses; National Metrology Laboratory (NML-CSIR), Pretoria, SOUTH AFRICA
Ruiz, Manuel; Industrial Technology Development Institute (ITDI), PHILIPPINES
Leones, Sabino Paulo B.; Industrial Technology Development Institute (ITDI), PHILIPPINES
Quelle/Jahr IEEE Transactions on Instrumentation and Measurement: 54 (2005), 2, 795 - 798
ISSN 0018-9456
DOI
Verlag Piscataway, NJ: IEEE
Konferenzangaben CPEM 2004, London, 27, June - 02, July, 2004, UK
Freie Schlagworte AC-DC transfer ; international comparison ; thermal voltage converter
Zusammenfassung This paper presents preliminary results of the APMP.EM-K6a key comparison of ac-dc transfer conducted during 2000 to 2003. The ac-dc voltage transfer difference of the travelling standard was measured at 3 V and selected frequencies from 1 kHz to 1 MHz that included those of the CCEM-K6a key comparison. Also discussed are the methods used to compare thermal voltage converters with different input connectors, and to link the results to the parent Consultative Committee on Electricity and Magnetism (CCEM) comparison.

Zitierung

Budovsky, I., Bin Zainal Abidin, A. R., Yan, A. Y. K., Liu, L., Rustagi, V. K., Govil, A. K., Klonz, M., Wei, Y.-C., Early, M. D., Sasaki, H., Fujiki, H., Kumtawee, C., Charoensook, A., Kwon, S.-W., Son, N. A., Suprianto, B., Temba, M., Ruiz, M., & Leones, S. P. B. (2005). APMP international comparison of ac-dc transfer standards at the lowest attainable level of uncertainty. IEEE Transactions on Instrumentation and Measurement, 54(2), 795–798. https://doi.org/10.1109/tim.2004.843088