| Zugriffsnummer | 17458 |
| Dokumenttyp | Konferenzartikel |
| Sprache | Englisch |
| Titel | Comparison measurements of MIKES-RAUTE 20 kN·m torque reference device with the PTB |
| Autor(in); Institution |
Pusa, Aimo; MIKES-RAUTE, Lahti, FINLAND
Röske, Dirk; 1.2, Festkörpermechanik, PTB-Braunschweig
Sachs, Michael; MIKES-RAUTE, Lathi, FINLAND
|
| Quelle/Jahr | 19th International Conference on Force, Mass and Torque, IMEKO TC 3: unrefereed full papers:(2005) |
| Availability | [CD-ROM] ; file name: 33.pdf |
| Herausgeber(in) |
Eltawil, Alaa eldin A.; National Institute of Standards (NIS), EGYPT
|
| URL | |
| Verlag | Giza: National Institute of Standards (NIS) |
| Konferenzangaben | 19th International Conference on Force, Mass & Torque. Theory and Application in Laboratories and Industries, Cairo, 19-23, February, 2005, Egypt |
| Freie Schlagworte | vertical torque reference calibration device ; torque calibration ; comparison measurements |
| Zusammenfassung | This article deals with the vertical 20 kN·m torque reference calibration device newly built by MIKES-RAUTE. It widens the possibilities of the MIKES-RAUTE Mass and Force Laboratory in the field of torque calibrations from 2 kN·m up to 20 kN·m. A brief overview about the construction and setup of the device is included in this report. Also some information about the applied calibration process is given. The main part of this article describes comparison measurements with the PTB, Germany, from summer 2004 and presents their results. They show that the device will attain an uncertainty level of 5·10-4 or better over the entire measurement range. |
Zitierung
Pusa, A., Röske, D., & Sachs, M. (2005). Comparison measurements of MIKES-RAUTE 20 kN·m torque reference device with the PTB. 19th International Conference on Force, Mass & Torque. Theory and Application in Laboratories and Industries, Cairo, 19-23, February, 2005, Egypt.