Zugriffsnummer 16804
Dokumenttyp Konferenzartikel
Sprache Englisch
Titel Thermodynamic temperature of the freezing point of copper measured by noise thermometry
Autor(in); Institution
Edler, Frank; 7.31, Thermometrie, PTB-Berlin
Tegeler, Erich; 7.31, Thermometrie, PTB-Berlin
Zimmermann, E.; Forschungszentrum Jülich GmbH, Jülich, GERMANY
Quelle/Jahr Temperature - its measurement and control in science and industry: 1 (2003), 13 - 18
Schriftenreihe AIP Conference Proceedings: 684
Herausgeber(in)
Ripple, Dean C.; American Institute of Physics
ISBN 0-7354-0153-5
DOI
Verlag Melville, NY: AIP
Konferenzangaben TEMPERATURE 2002, The 8th Symposium: Its Measurement and Control in Science and Industry, Chicago, Illinois, 21-24, October, 2002, USA
Klassifikationscode PACS: 07.20.Dt ; PACS: 64.70.Dv
Freie Schlagworte temperature measurement ; freezing ; copper ; thermometers ; resistors ; thermal noise ; thermodynamic properties
Zusammenfassung The thermodynamic temperature of the freezing point of copper was measured by noise thermometric methods. A conventional approach was used which combines the noise voltage of the measuring resistor with that of a reference resistor maintained at a known temperature. Besides this comparison method the thermometer is characterized by a two-channel arrangement to eliminate parasitic noise of electronic components by cross correlation. The thermo-dynamic temperature measured at the freezing point of copper amounts to a value of 1084.54 °C ± 0.12 K (k = 2). This corresponds to the value of the ITS-90 temperature of this fixed point of 1084.62 °C within its thermodynamic uncertainty of 0.12 K (k = 2).

Zitierung

Edler, F., Tegeler, E., & Zimmermann, E. (2003). Thermodynamic temperature of the freezing point of copper measured by noise thermometry. TEMPERATURE 2002, The 8th Symposium: Its Measurement and Control in Science and Industry, Chicago, Illinois, 21-24, October, 2002, USA. https://doi.org/10.1063/1.1627093