| Zugriffsnummer | 16286 |
| Dokumenttyp | Zeitschriftenartikel |
| Sprache | Englisch |
| Titel | Amplitude and phase calibration of hydrophones by heterodyne and time-gated time-delay spectrometry |
| Autor(in); Institution |
Koch, Christian; 1.43, Ultraschall, PTB-Braunschweig
|
| Quelle/Jahr | IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control: 50 (2003), 3, 344 - 348 |
| ISSN | 0885-3010 |
| Verlag | New York, NY: IEEE |
| Zusammenfassung | Two methods for the amplitude and phase calibration of hydrophones based on time-delay spectrometry are presented. A time gating technique and heterodyne frequency conversion allows the phase-locked, i. e. coherent detection of hydrophone signals. Amplitude and phase of the hydrophone sensitivity are obtained in the frequency range 1 - 20 MHz. In addition, to show the broad-band capabilities of the methods, amplitude values are measured between 4 and 50 MHz. |