Zugriffsnummer 16026
Dokumenttyp Konferenzartikel in Zeitschrift
Sprache Englisch
Titel A traceable precision current source for currents between 100 aA and 10 pA
Autor(in); Institution
Willenberg, Gerd-Dietmar; 2.12, Gleichstromgrößen, PTB-Braunschweig
Tauscher, Heinz-Norbert; 2.12, Gleichstromgrößen, PTB-Braunschweig
Warnecke, Peter; 2.12, Gleichstromgrößen, PTB-Braunschweig
Quelle/Jahr IEEE Transactions on Instrumentation and Measurement: 52 (2003), 2, 436 - 439
ISSN 0018-9456
Verlag New York, NY: IEEE
Konferenzangaben Conference on Precision Electromagnetic Measurements ; CPEM 2002, Ottawa, 16-21, June, 2002, Canada
Freie Schlagworte Calibration ; current ; picoamperemeter
Zusammenfassung A new calibration setup for picoamperemeters is presented. Owing to the use of a novel voltage ramp generator with increased linearity, the useful measuring time is increased to about 35 min, which reduces the calibration uncertainty of picoamperemeters with long time constants.

Zitierung

Willenberg, G.-D., Tauscher, H.-N., & Warnecke, P. (2003). A traceable precision current source for currents between 100 aA and 10 pA. IEEE Transactions on Instrumentation and Measurement, 52(2), 436–439.