| Zugriffsnummer | 14906 |
| Dokumenttyp | Konferenzartikel |
| Sprache | Englisch |
| Titel | Inductance calibration in the frequency range from 50 Hz to 1 MHz at PTB [poster] |
| Autor(in); Institution |
Hanke, Rüdiger; 2.001, Qualitätsmanagement und Messunsicherheit, PTB-Braunschweig
Kölling, Axel; 2.11, Wechselstromgrößen, PTB-Braunschweig
Melcher, Jürgen; 2.11, Wechselstromgrößen, PTB-Braunschweig
|
| Quelle/Jahr | Conference digest : 2002 Conference on Precision Electromagnetic Measurements:(2002), 186 - 187 |
| ISBN | 0-7803-7242-5 |
| Verlag | Piscataway, NJ: IEEE |
| Konferenzangaben | Conference on Precision Electromagnetic Measurements ; CPEM 2002, Ottawa, 16-21, June, 2002, Canada |
| Zusammenfassung | Inductance calibrations at PTB are carried out with a Maxwell-Wien bridge. This bridge has the advantage that the first order bridge equation is independent of frequency which means that the bridge has the potential to be operated in a large frequency range. However, the actual frequency range, from power frequencies up to 1 MHz, requires an investigation of higher order effects. |