Zugriffsnummer 14822
Dokumenttyp Konferenzartikel in Zeitschrift
Sprache Englisch
Titel Transfer of the efficiency calibration of Germanium gamma-ray detectors using the GESPECOR software
Autor(in); Institution
Sima, Octavian; Bucharest University, Physics Department, Bucharest, ROMANIA
Arnold, Dirk; 6.12, Umweltradioaktivität, PTB-Braunschweig
Quelle/Jahr Proceedings of the conference on radionuclide metrology and its applications, ICRM 2001. Applied radiation and isotopes: 56 (2002), 1-2, 71 - 75
Herausgeber(in)
Coursey, B. M.; National Institute of Standards and Technology, Physics Laboratory, Ionizing Radiation Division, Gaithersburg, Md., USA
Janßen, Herbert; 6.1, Radioaktivität, PTB-Braunschweig
ISSN 0969-8043
Verlag Oxford [u.a.]: Pergamon
Konferenzangaben 13th International Conference on Radionuclide Metrology and its Applications ; ICRM 2001, Braunschweig, 14-18, Mai, 2001, Germany
Freie Schlagworte gamma-spectrometry ; efficiency transfer ; matrix effects ; geometry effects ; GESPECOR
Zusammenfassung The GESPECOR software was extended to incorporate procedures for the computation of the efficiency transfer factor for cases of practical interest:(a)sources with identical geometry, but different matrices; (b) sources with similar(but not identical)geometry;(c)transfer from a point source to a volume source. Fast and accurate algorithms, based on correlated sampling, were implemented for solving the first two cases. A procedure to take into account the imperfect charge collection in the detector was implemented.

Zitierung

Sima, O. & Arnold, D. (2002). Transfer of the efficiency calibration of Germanium gamma-ray detectors using the GESPECOR software. Applied radiation and isotopes, 56(1-2), 71–75.

Exportieren