Zugriffsnummer 13903
Dokumenttyp Konferenzartikel
Sprache Englisch
Titel Calibration facility for infrared thermography systems at PTB
Autor(in); Institution
Hartmann, Jürgen; 7.21, Temperaturstrahlung, PTB-Berlin
Gutschwager, Berndt; 7.21, Temperaturstrahlung, PTB-Berlin
Fischer, Joachim; 7.21, Temperaturstrahlung, PTB-Berlin
Quelle/Jahr Sensor 2001 : 10th international conference, May 8 - 10, 2001, Exhibition Centre Nuremberg, Germany ; proceedings: 2 (2001), 233 - 238
Verlag Wunstorf: AMA Service GmbH
Konferenzangaben 10th International Conference "Sensor 2001", Nuremberg, 8-10, May, 2001, Germany
Zusammenfassung Physikalisch-Technische Bundesanstalt (PTB) has set-up a measurement facility for calibration of thermography systems based on high quality blackbody radiators covering the temperature range from -60°C up to 3000°C. The traceability to ITS-90 is either performed by platinum resistance thermometers (below 962°C) or by radiation thermometry with respect to the gold fixed-point (above 962°C). Additionally, the facility enables investigation of the angular and temperature resolution of the imaging system. A collimator system is used for enabling investigation at infinite object distance. Different environmental conditions can be reproduced by putting the thermography systems in a climate chamber, allowing detailed controlling of humidity and temperature.

Zitierung

Hartmann, J., Gutschwager, B., & Fischer, J. (2001). Calibration facility for infrared thermography systems at PTB. 10th International Conference "Sensor 2001", Nuremberg, 8-10, May, 2001, Germany.