Zugriffsnummer 13697
Dokumenttyp Konferenzartikel
Sprache Englisch
Titel Six nanoradian in 2π radian - a primary standard for angle measurement [poster]
Autor(in); Institution
Probst, Reinhard; 5.202, Winkelmesstechnik, PTB-Braunschweig
Krause, Michael; 5.202, Winkelmesstechnik, PTB-Braunschweig
Quelle/Jahr Precision engineering, nanotechnology : proceedings of the Euspen 2nd International Conference, European Society for Precision Engineering and Nanotechnology. Vol. 1:(2001), 326 - 329
Herausgeber(in)
Balsamo, A.; CNR-IMGC, Torino, ITALY
Verlag Turin:
Konferenzangaben Euspen 2nd International Conference, Torino, 27-31, May, 2001, Italy
Freie Schlagworte angle measurement ; angle standard ; angle comparator ; angle calibration ; Fourier analysis
Zusammenfassung A self-calibration method based on the Fourier analysis is described and applied to the divided circle of an angle comparator. Results are presented and compared with a second calibration method. An uncertainty of measurement U = 23 nrad or 0,0047'' (k = 2) is evaluated for the comparator.

Zitierung

Probst, R. & Krause, M. (2001). Six nanoradian in 2π radian - a primary standard for angle measurement [poster]. Euspen 2nd International Conference, Torino, 27-31, May, 2001, Italy.

Exportieren

PTB-Publica Menü

Sprache wechseln: uk flag