Zugriffsnummer 13618
Dokumenttyp Zeitschriftenartikel
Sprache Englisch
Titel Half-life measurements of Europium radionuclides and the long-term stability of detectors
Autor(in); Institution
Siegert, Helmut; 6.11, Aktivitätseinheit und Messverfahren, PTB-Braunschweig
Schrader, Heinrich; 6.11, Aktivitätseinheit und Messverfahren, PTB-Braunschweig
Schötzig, Ulrich; 6.11, Aktivitätseinheit und Messverfahren, PTB-Braunschweig
Quelle/Jahr Applied Radiation and Isotopes: 49 (1998), 9-11, 1397 - 1401
ISSN 0969-8043
Verlag Amsterdam: Elsevier
Zusammenfassung Consideration of the long-term stability of detectors has proved to be very important in half-life measurements. For example, studies of a 226RA reference source with a pressurised 4πγ ionization chamber for stability checks showed an average increase in efficiency of 0.005% per year. Taking this effect into account, half-lives of T½(152Eu) = 4936.6(20)d and T½(154Eu) = 3138.1(16)d were determined. Using the half-life value of 152Eu as a reference, the time-dependent decrease of the efficiency of semiconductor detectors at different photon energies was determined for periods of the order of twenty years. Values decreased by an average of less than 0.01% per year for a high-purity Ge detector, by about 0.03% per year for a GE(Li)detector and by about 0.07% per year for a SI(Li)detector. Half-life values of T½(154Eu) = 3146.(11)d and T½(155Eu) = 1739(8)d were determined. The uncertainty of the 155Eu half-life value was reduced by a factor of three in comparison to an earlier PTB measurement to which this correction had not been applied.

Zitierung

Siegert, H., Schrader, H., & Schötzig, U. (1998). Half-life measurements of Europium radionuclides and the long-term stability of detectors. Applied Radiation and Isotopes, 49(9-11), 1397–1401.

Exportieren

PTB-Publica Menü