Zugriffsnummer 12972
Dokumenttyp Konferenzartikel
Sprache Englisch
Titel Absolute calibration of ACIS X-ray CCDs using calculable, undispersed synchrotron radiation
Autor(in); Institution
Bautz, M. W.; MIT Center for Space Research, Cambridge, USA
Pivovaroff, M. J.; Space Research, Cambridge, USA
Kissel, S. E.; Space Research, Cambridge, USA
Prigozhin, G. Y.; Space Research, Cambridge, USA
Isobe, T.; Harvard-Smithsonian CfA/CXO, Cambridge, USA
Jones, S. E.; Space Research, Cambridge, USA
Ricker, G. R.; Space Research, Cambridge, USA
Thornagel, Reiner; 7.12, Röntgenradiometrie, PTB-Berlin
Kraft, Stefan; 7.12, Röntgenradiometrie, PTB-Berlin
Scholze, Frank; 7.12, Röntgenradiometrie, PTB-Berlin
Ulm, Gerhard; 7.1, Photonenradiometrie, PTB-Berlin
Quelle/Jahr X-ray optics, instruments and missions III:(2000), 53 - 67
Schriftenreihe Proceedings of SPIE: 4012
Herausgeber(in)
Trümper, Joachim E.
Aschenbach, Bernd
ISSN 0277-786X
ISBN 0-8194-3637-2
DOI
Verlag Washington: SPIE
Konferenzangaben X-Ray Optics, Instruments, and Missions III, Munich, 27 - 29 March 2000, Germany
Freie Schlagworte CCDs ; X-rays ; ACIS ; Chandra
Zusammenfassung We have performed a series of measurements with the Physikalisch-Technische Bundesanstalt beamline of the electron storage ring BESSY I which provide the basis for the absolute calibration of the Advanced CCD Imaging Spectrometer (ACIS). ACIS is a prime focal plane instrument aboard the recently-launched Chandra X-ray Observatory. We have achieved an absolute detection efficiency knowledge accurate to better than 5% over the 0.3 - 4 keV band. We describe our measurement and analysis techniques, including our detector response modeling and pileup corrections. We summarize a variety of external and internal consistency checks which provide the basis for our error estimates. We discuss the factors limiting the accuracy of our measurements.

Zitierung

Bautz, M. W., Pivovaroff, M. J., Kissel, S. E., Prigozhin, G. Y., Isobe, T., Jones, S. E., Ricker, G. R., Thornagel, R., Kraft, S., Scholze, F., & Ulm, G. (2000). Absolute calibration of ACIS X-ray CCDs using calculable, undispersed synchrotron radiation. X-Ray Optics, Instruments, and Missions III, Munich, 27 - 29 March 2000, Germany. https://doi.org/10.1117/12.391582