Zugriffsnummer 12968
Dokumenttyp Konferenzartikel
Sprache Englisch
Titel Distributed read-out imaging devices for X-ray imaging spectroscopy
Autor(in); Institution
den Hartog, Roland; Astrophysics Division, Noordwijk, THE NETHERLANDS
Martin, D.; Astrophysics Division, Noordwijk, THE NETHERLANDS
Kozorezov, A.; School of Chemistry and Physics, Lancaster, UK
Verhoeve, P.; Astrophysics Division, Noordwijk, THE NETHERLANDS
Rando, N.; Astrophysics Division, Noordwijk, THE NETHERLANDS
Peacock, A.; Astrophysics Division, Noordwijk, THE NETHERLANDS
Brammertz, G.; Astrophysics Division, Noordwijk, THE NETHERLANDS
Krumrey, Michael; 7.12, Röntgenradiometrie, PTB-Berlin
Goldie, D. J.; Oxford Instruments Scientific Research Division, Cambridge, UK
Venn, R.; Cambridge MicroFab Ltd., Cambridge, UK
Quelle/Jahr X-ray optics, instruments and missions III:(2000), 237 - 248
Schriftenreihe Proceedings of SPIE: 4012
Herausgeber(in)
Trümper, Joachim
Aschenbach, Bernd
ISSN 0277-786X
ISBN 0-8194-3637-2
DOI
Verlag Bellingham, Wash.: SPIE
Konferenzangaben Conference on X-ray Optics, Instruments, and Missions III, Munich, 27-29, March, 2000, Germany
Freie Schlagworte Superconducting tunnel junctions ; X-ray detectors ; X-ray imaging ; X-rax spectroscopy ; DROID ; XEUS
Zusammenfassung We present an experimental study of the performance of Distributed Read-Out Imaging Devices (DROIDs), 1- and 2-dimensional photon-counting imaging spectrometers, based on Ta/Al-based STJs placed on a Ta absorber. Results obtained with highly collimated illumination with 10 keV X-ray photons clearly demonstrate the imaging capabilities of 2-dimensional DROIDs. The derived spatial FWHM resolution is 7 µm for a 200 × 200 μm² absorber. With a 1-D DROID we have measured an intrinsic energy resolution of 15 eV FWHM for 6 keV photons. At high energies (E ) 6 keV) the resolution is limited by spatial fluctuations in the qp recombination rate.

Zitierung

den Hartog, R., Martin, D., Kozorezov, A., Verhoeve, P., Rando, N., Peacock, A., Brammertz, G., Krumrey, M., Goldie, D. J., & Venn, R. (2000). Distributed read-out imaging devices for X-ray imaging spectroscopy. Conference on X-ray Optics, Instruments, and Missions III, Munich, 27-29, March, 2000, Germany. https://doi.org/10.1117/12.391559

Exportieren

PTB-Publica Menü