Zugriffsnummer 12964
Dokumenttyp Zeitschriftenartikel
Sprache Englisch
Titel The laboratory calibration of the SOHO coronal diagnostic spectrometer
Autor(in); Institution
Lang, J.; Rutherford Appleton Laboratory, Chilton, UK
Kent, B. J.; Rutherford Appleton Laboratory, Chilton, UK
Breeveld, A. A.; Mullard Space Science Laboratory, Holmbury St Mary, UK
Breeveld, E. R.; Mullard Space Science Laboratory, Holmbury St Mary, UK
Bromages, B. J. I.; University of Central Lancashire, Preston, UK
Hollandt, Jörg; 7.11, UV- und VUV-Radiometrie, PTB-Berlin
Payne, J.; Rutherford Appleton Laboratory, Chilton, UK
Pike, C. D.; Rutherford Appleton Laboratory, Chilton, UK
Thompson, W. T.; SM&A Corporation, Largo, USA
Quelle/Jahr Journal of Optics A: 2 (2000), 2, 88 - 106
ISSN 1464-4258
DOI
Verlag Bristol: Inst. of Physics Publ.
Klassifikationscode PACS: 95.55.Ev ; PACS: 96.60.Tf ; PACS: 06.20.F- ; PACS: 96.60.P- ; PACS: 07.60.Rd ; PACS: 07.60.Dq
Freie Schlagworte Solar spectroscopy ; extreme UV spectroscopy ; radiometry ; SOHO
Zusammenfassung The laboratory end-to-end testing of the Coronal Diagnostic Spectrometer (CDS) experiment on the ESA/NASA SOHO mission is reported. A brief overview of CDS, which operates in the extreme-ultraviolet wavelength range, is given. Pertinent details of the calibration source are presented, followed by an account of the source beam characterization. A section is devoted to the determination of the instrument apertures and this includes an outline of the measurements, the results from both the grazing incidence and normal incidence aperture scans and their interpretation to yield estimates aperture areas. Next the measurement of spectrometer bandwidths and their comparison with expected values are described. Then the pre-launch wavelength calibrations are obtained. The section on sensitivity starts with an evaluation of the effects of polarization on the measurements. The expected sensitivities are then derived. The measurements, their analysis and a comparison of measured and expected sensitivities are presented for both the normal incidence and grazing incidence spectrometers. The application of the laboratory calibration to in-flight solar data is discussed.