Zugriffsnummer 12705
Dokumenttyp Konferenzartikel
Sprache Englisch
Titel The influence of misalignment on torque transducers
Autor(in); Institution
Brüge, Andreas; 1.13, Drehmoment, PTB-Braunschweig
Peschel, Diedert; 1.13, Drehmoment, PTB-Braunschweig
Röske, Dirk; 1.13, Drehmoment, PTB-Braunschweig
Quelle/Jahr Proceedings, XVI IMEKO World Congress, IMEKO 2000: 3 (2000), 13 - 17
Herausgeber(in)
Osanna, P. H.; Austrian Society for Measurement and Automation
ISBN 3-901888-05-5
Verlag Wien: Eigenverlag ÖGMA
Konferenzangaben XVI IMEKO World Congress ; IMEKO 2000, Vienna, 25-28, September, 2000, Austria
Freie Schlagworte torque ; calibration ; misalignment ; flexible coupling
Zusammenfassung Misalignment during the mounting of torque transducers in calibration devices can cause parasitic bending inside the transducers. Flexible couplings allow the contribution of misalignment to the measurement uncertainty to be reduced to a negligible amount. The proceedings are also available on CD-ROM.

Zitierung

Brüge, A., Peschel, D., & Röske, D. (2000). The influence of misalignment on torque transducers. XVI IMEKO World Congress ; IMEKO 2000, Vienna, 25-28, September, 2000, Austria.