Titel Internal and external comparison measurements for ensuring accuracy and traceability in flowmeter calibration
Autor(in); Institution
Pöschel, Walter; 1.32, Flüssigkeitsmessgeräte, PTB-Braunschweig
Quelle/Jahr Fluid flow measurement : 4th international symposium, Hyatt Regency Denver, June 27 - 30, 1999, Denver, Colorado, USA ; proceedings:(1999), 13 S.
Availability [CD-ROM] ; file name: 15_03.pdf
Herausgeber(in)
North American Fluid Flow Measurement Council
Verlag Denver, Colo.:
Konferenzangaben 4th International Symposium on Fluid Flow Measurement, Denver, Colorado, 27-30, June, 1999, USA
Freie Schlagworte liquid flow measurement ; primary standard ; accuracy ; intercomparison traceability
Zusammenfassung Comparison measurements become more important to perform traceability an to confirm the accuracy of flowmeter calibration facilities. But intercomparsisons between different laboratories are very complicated, expensive and time consuming. Therefor a method is proposed to prove different components of the total measurement uncertainty by internal comparison measurements between different flow measurement devices within the laboratory. The paper describes this method and it's realization in the new primary standard for water flow measurement at PTB (being under construction now).
Rechteinformation 1999, North American Fluid Flow Measurement Council

Zitierung

Pöschel, W. (1999). Internal and external comparison measurements for ensuring accuracy and traceability in flowmeter calibration. 4th International Symposium on Fluid Flow Measurement, Denver, Colorado, 27-30, June, 1999, USA.