| Titel | New approach to calibrating conductivity meters in the low conductivity range |
| Autor(in); Institution |
Spitzer, Petra; 3.1, Metrologie in der Chemie, PTB-Braunschweig
Rossi, Bruno; Lab Water Division, Millipore, St. Quentin-Yvelines, FRANCE
Gaignet, Yves; Lab Water Division, Millipore, St. Quentin-Yvelines, FRANCE
Mabic, Stéphane; Lab Water Division, Millipore, St. Quentin-Yvelines, FRANCE
Sudmeier, Uwe; 3.1, Metrologie in der Chemie, PTB-Braunschweig
|
| Quelle/Jahr | Accreditation and Quality Assurance: 10 (2005), 3, 78 - 81 |
| ISSN | 0949-1775 (PRINT) ; 1432-0517 (ONLINE) |
| DOI | |
| Verlag | Berlin [u.a.]: Springer |
| Freie Schlagworte | Electrolytic conductivity ; High purity water ; Calibration ; Primary methods ; Metrology in chemistry |
| Zusammenfassung | There is currently a major issue with the calibration of conductivity meters used for high purity water: the lack of availability of a reference material or reference methods for low conductivity ranges (conductivity below 1 μS cm-1 at 25.0 °C, resistivity >1 MΩ cm at 25.0 °C). This paper describes the current status of conductivity measurements in high purity water. A new and improved approach, currently being investigated, should allow us to make the calibration of conductivity meters used for low conductivity ranges traceable to the SI. |