Autor(in); Institution
Weyand, Kurt; 2.5, Halbleiterphysik und Magnetismus, PTB-Braunschweig
Ketzler, Rainer; 2.5, Halbleiterphysik und Magnetismus, PTB-Braunschweig
Kitschke, Manfred; 2.5, Halbleiterphysik und Magnetismus, PTB-Braunschweig
Xu, Jiheng; shanghai University, Faculty of Automation, Shanghai, CHINA
Quelle/Jahr 2004 Conference on Precision Electromagnetic Measurements Digest:(2004), 168 - 169
ISBN 0-7803-8493-8
Berichtsnummer 04CH37570
Verlag Piscataway, NJ: IEEE
Konferenzangaben 2004 Conference on Precision Electromagnetic Measurements, CPEM, London, 27, June - 02, July, 2004, UK
Zusammenfassung For the calibration of magnetic instruments a dc current source controlled by Nuclear Magnetic Resonance (NMR) has been developed for currents in the range from 0.3 to 1 A. Depending on the absorption line-width of the NMR sample a resolution of less than 1 μA and a current instability of less than 1 ppm over 8 hours can be achieved.

Zitierung

Weyand, K., Ketzler, R., Kitschke, M., & Xu, J. (2004). DC-current source controlled by NMR. 2004 Conference on Precision Electromagnetic Measurements, CPEM, London, 27, June - 02, July, 2004, UK.